Optical Design

The source material for these pages is the CDR optical chapter, the February 2026 post-CoDR optical update, and my notes/memory The post-CoDR update controls the design intent for the ZSpec channel split and detector format. -Jeb

A key todo is to revise the requirement/design language for compatibility with this agreed upon updated R.

Early PD Optical Design Update.

Figure placeholders for this page

  • One clean optical block diagram showing RNAS → ZFront → ZSpec/ZImager (& R/L inst)

  • One common-coordinate sketch showing guide field, ZImager field, ZSpec slit center, and calibration field.

The present optical concept is composed of into four key parts:

Optical assemblies

Assembly

Function

Current design intent

RNAS interface

Receive the Keck I beam and preserve future AO/guide compatibility.

ZShooter sits on-axis at K1 RNAS, where the deployable tertiary can feed it quickly. The mechanical/optical interface must leave volume for the facility guide and future AO wavefront-sensing hardware.

ZFront

Select science/calibration paths and send light to ZSpec, ZImager, or neighboring RNAS instruments.

Two-stage light-path selection, with a calibration launch upstream of the selection.

ZSpec

Six science channels of fixed-format echelle spectroscopy.

Two spectrometers: one visible and one infrared. Each spectrometer is split into three detector channels formatted on 2k × 2k active areas.

ZImager

Simultaneous tri-band imaging for photometric anchoring, high-cadence work, imaging-only fallback science, and acquisition support.

Refractive collimator, dichroic split into u / selectable / z-like channels, and one fast qCMOS detector module per channel.

RNAS Focus

ZShooter is allocated to the Keck I Right Nasmyth platform. The native Nasmyth focus is described as approximately 1.5 m above the platform floor and approximately 0.39 m outside the elevation bearing. This motivates the use of front-end relay and selection optics: the focus is too close to the bearing to let the full spectrograph architecture simply grow around the native focal plane without imposing significant packaging constraints.

The optical design assumes that ZShooter is a natural-seeing instrument capable of utilizing exceptional seeing. The future K1 AO STRATA system is treated as a performance enhancer increasing the number of “exceptional seeing” nights rather than as an enabling condition. ZShooter will be ready to exploit narrower delivered PSFs for reduced sky background and higher spectral resolution without concomitant slit losses when AO becomes available.

Figure placeholder: K1 RNAS nominal focus geometry from cad/drawings/k1rnas.pdf or the equivalent released CAD/ICD view. This should show the K1 elevation bearing, nominal telescope focus, the ZImager fold path, and the straight-through ZSpec path.

Confirm the allowable secondary-z focus adjustment range, subject to some nominal induced aberration/vignetting penalty, and then what additional shift STRATA/K1AOF can practically deliver for L/R instruments without (or with quantified, minimal, i.e. fold mirror sizes) changes to the ZFront relay assumptions.

ZFront

ZFront is the telescope-facing beam switchyard and calibration assembly. It selects where the telescope beam or calibration beam goes.

The current concept is a two-stage light-path selector. The first selector chooses whether the incoming field goes directly to ZImager, to ZImager with the central field passed on to ZSpec, or to the second selector. The second selector chooses the ZSpec dichroic/pre-optics path or a feed to an instrument to the left or right of ZShooter on the RNAS platform. This implementation naturally allows the central science field to be passed to neighboring instruments as well while imaging the surrounding with ZImager, but maintaining such capability is not a ZShooter science requirement.

The calibration launch is planned upstream of the first selector so that a common calibration reference can be delivered to both systems. The baseline concept is a fiber-fed integrating sphere that illuminates a roughly 20–30 arcsec central field, with lamp sources for wavelength calibration and continuum/flat-field work. Hollow-cathode lamps such as ThAr, ThArNe, ThNe, or UNe, together with LED and quartz sources: the specific sources remain under evaluation. A point-source mask is also attractive because it would let the same calibration path probe ZSpec/ZImager co-alignment and PSF behavior rather than only flat-field response.

Current optical review queue

  • ZFront calibration path: define whether the point-source/pinhole mode is a requirement or an AIT convenience.

ZSpec Summary

For details, see ZSpec.

ZSpec is best described as a six-channel, two-spectrometer, fixed-format echelle. The visible spectrometer is split into Blue, Green, and Red channels; the infrared spectrometer is split into YJ, H, and K channels. This preserves the broad 310–2450 nm science grasp while reducing the number of slits and the echelle gratings, avoiding detector and echelle mosaics, and reducing the previous large anamorphic excursions of the conceptual design while delivering improved optical efficiency.

Spectrometer

Channel

Approximate passband

Design rationale

Visible

Blue

308–420 nm

Highest leverage for atmospheric-cutoff science, hot stars, D/H, and flash-ionization diagnostics and partially isolates the region with the most difficult coatings, glass transmission, and detector QE.

Visible

Green

400–600 nm

Places the highest-throughput optical region in its own channel rather than on the roll-off of a broader blue/red cross-disperser design.

Visible

Red

580–980 nm

Carries the red optical diagnostics and bridges toward the YJ channel without forcing one detector/coating solution across the full visible range. Covers Hα through Ca triplet and red continuum/absorption diagnostics.

Infrared

YJ

950–1350 nm

Avoids an IR detector mosaic and isolates the lowest-background NIR range.

Infrared

H

1457–1848 nm

Improves the efficiency balance relative to the conceptual studies YJ/HK split.

Infrared

K

1972–2482 nm

Contains the thermal-background-sensitive portion of the design in one channel.

Jason, I’m inclined to axe the rationale above or qualify it here. To my mind this is driven by detector area demands and blaze efficiencies given R and m needs with pupil size.

ZImager Summary

ZImager is a co-aligned tri-band imager. It supports four roles: pre/post-imaging around ZSpec observations, photometric anchoring of spectra, imaging-only fallback for targets beyond spectroscopic reach, and millisecond-to-second cadence observations of compact and rapidly variable sources. It is also useful operationally because it can help identify faint transients, moving targets, host-galaxy context, and slit-placement ambiguities before committing ZSpec exposure time.

It’s optical layout uses a refractive collimator over a 1–3 arcmin field, two dichroics and a fold mirror to create three channels with fast cameras feeding one ORCA-Quest qCMOS detector per band. The nominal filters are u, one selectable middle band, and z. The middle channel is expected to include both standard broadband options and may carry narrow filters where science demand and throughput justify them.

For details, see ZImager.

Guiding and Wavefront Sensing

Routine telescope guiding is expected to be handled by a facility guide system developed for K1 RNAS, with ZShooter as the initial client. The same facility work is expected to preserve the volume and optical interfaces needed for natural-guide-star and sodium-laser wavefront sensing associated with the Keck I AO facility. This keeps ZShooter from carrying a one-off guider solution that would later fight the AO architecture or be replicated by neighboring instruments.

The key ZShooter optical implications are straightforward. First, the slit, ZImager field, guide field, and future WFS references must be tied to a stable and well-documented coordinate model. Second, acquisition support must work before AO is available. Third, once seeing enhancement is available, the optical design should turn a smaller delivered PSF into better S/N at higher spectral resolution without requiring a redesign of ZSpec.

A detailed white-paper on the relevant coordinate transforms is available here `zs_cood_math`_.

Calibration and Co-alignment

The optical calibration strategy is meant to be repeatable and low overhead. ZFront provides common internal calibration delivery for a central common field; ZImager supplies field photometry and WCS context; ZSpec supplies fixed-format spectra whose stability can be tracked against arcs, flats, standards, and sky features. The system should therefore be aligned and documented as an integrated optical instrument, not as a spectrograph plus an unrelated camera.

At minimum, the optical design and AIT plan will verify:

  • ZSpec/ZImager common-field registration and its repeatability across light-path selector moves

  • Slit-center, ZImager channel WCS, and guide-system reference consistency, including reregistration procedures.

  • Wavelength calibration stability versus slit-selector repeatability, rotator angle, thermal state, and time.

  • ZImager photometric flatness and color terms for the selected broadband filters.

  • Calibration-beam uniformity and its ability to support both flat-fielding and co-alignment checks.